Exposure Mode: Manual
Technique Factor Range: 55kV and 70Kv one step footswitch control
Accuracy of technique factors: mA +/- 10% of pre-indicated value kVP +/- 10% of pre-indicated value Time +/- 0.025s
Generator output: Measured in volts by non-evasive kVP Meter
Tube Current: Direct measurement at the cathode of the x-ray tube.
Beam Current: 250uA fixed
Focal Spot: 60 microns
Clear Opening: 25.4cm(10.0)
Throat Depth: 30.5cm(12.0)
Field of View: 75mm (3.0)
Image Display: USB digital conversion to PC or Laptop C-Thru Image save basic software, standard
Image Enhance: Hi / Low frame, Image Flip, Image Negate, 2X Contrast, Image Brightness Control
Stand / Articulating Arm
Stand footprint: 22" x 24" x 48",
C-Arm Weight: 16 lbs.
Stand Weight: 160 lbs.
Dimensions:
17L x 17W x 1.75H
Software:
Stand Alone

Main Power On / Off Switch
On / Off C-Arm Keyswitch
Image Negate Switch
Noise Reduce Switch
Hi / Low frame Switch
Image Flip Switch
2X Contrast Switch
Image Brightness Control Knob

AC Voltage:
115 VAC , 50 / 60 HZ
Technique Display:
KV, uA, Accumulated Time, AKR, Accumulated AKR, X-Ray On / Off, Image Hold
DISCLAIMER
This item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If the item is subject to FDA regulation, we will verify your status as an authorized purchaser of this item before shipping of the item. If you have questions regarding such, please consult the FDA's WebSite for more information. If you are not a licensed practitioner, the purchase of this item may require a prescription.
Please contact us in advance for clarification."
This item has been intended for use in industrial and veterinarian applications only, and not for human medical use.
There is no 510k premarket approval on this item.

Warranty Information
1 Year Limited Warranty
C-Arm Specs
Console Specs
cthruweb_2016-2-1-1001003.png cthruweb_2016-2-1-1001003.png
Home
Contact
cthruweb_2016-2-1-1001003.png
750P
CTHRU 750SA
Copyright 2016 C-Thru Imaging LLC